Operator’s Manual
26
THERAPY SETTINGS (Continued)
3.
Pre-Condition Water
The
Pre-Condition Water
feature cools or warms the
water to a set temperature prior to starting patient therapy.
Use the Pre-Condition Water parameter to either disable
the Pre-Condition Water feature or set the pre-conditioned
water to a selected temperature. If a Pre-Condition Water
temperature is selected, then water pre-conditioning will
automatically start after entering the therapy screen.
Preconditioning will automatically end when the system
detects a valid patient temperature or that a pad has been
attached. This feature is useful to reduce the time needed
to cool or rewarm a patient.
•
From the
Normothermia Settings
screen or the
Hypothermia Settings
screen, press the button to the
right of the Pre-Condition Water parameter.
•
Use the Up and Down arrows to either disable the feature
or select the pre-condition water temperature.
Range:
Disable; 4°C to 40°C
39.2°F to 104°F
Increments:
1°C/°F
•
Press the
Save
button to save the new setting and close
the window.
•
Press the
Cancel
button to close the window without
saving the new settings.
4. Manual Control
•
From the
Normothermia Settings
screen or the
Hypothermia Settings
screen Press the button to the
right of the Manual Control Settings parameter.
Adjust to the desired temperature
•
Use the Up and Down arrows to either disable the feature
or select the Manual Control default temperature. The
selected temperature will be the default temperature
displayed in the Manual Control pop-up window.
Range:
Disable; 4°C to 40°C
39.2°F to 104.0°F
Increments:
1°C/°F
Adjust the desired time
•
Use the Up and Down arrows to select the Manual Control
maximum time duration. The selected time will be the
default time displayed in the Manual Control pop-up
window.
Range:
15 minutes to 99 hours
Increments: 15 minutes (from 0 to 8 hours)
1 hour (from 8 to 99 hours)
•
Press the
Start
button to save the new setting and close
the window.
•
Press the
Cancel
button to close the window without saving
the new settings.
5.
High Water Limit
Use the
High Water Limit
to set the maximum water
temperature delivered to the pads during therapy.
Reducing the high water temperature limit may be
appropriate for patients with fragile skin or other medical
conditions.
•
From the
Normothermia Settings
screen or the
Hypothermia Settings
screen, press the button to the
right of the High Water Limit parameter.
•
Use the Up and Down arrows to select the maximum
water temperature.
Range:
36°C to 40°C
97°F to 104°F
Increments: 1°C/°F
•
Press the
Save
button to save the new setting and close
the window.
•
Press the
Cancel
button to close the window without
saving the new settings.
NOTE:
The
High Water Limit
setting does not affect the
Manual Control
water temperature.
6. Low Water Limit
Use the
Low Water Limit
to set the minimum water
temperature delivered to the pads during therapy.
Increasing the low water temperature limit may be
appropriate for patients with fragile skin or other medical
conditions.
•
From the
Normothermia Settings
screen or the
Hypothermia Settings
screen, press the button to the
right of the Low Water Limit parameter.
•
Use the Up and Down arrows to select the minimum water
temperature.
Range:
4°C to 25°C
39°F to 77°F
Increments: 1°C/°F
•
Press the
Save
button to save the new setting and close
the window.
•
Press the
Cancel
button to close the window without
saving the new settings.
NOTE:
The
Low Water Limit
setting does not affect the
Manual Control
water temperature.
Содержание BD Arctic Sun 6000
Страница 1: ...14868...
Страница 2: ...Operator s Manual 2 tube...
Страница 57: ...The current Operator s Manual can be found at www medivance com manuals English 57 PK2800300 MM YYY...