Alarm
Problem
Solution
10 Patient
Temperature 1 Low
Medium Priority
Alarm
The Patient Temperature 1 reading
is below 31°C (87.8°F), and the
water temperature is below 31°C
(87.8°F), and the system is
continuing to cool the patient when
the system is in a patient control
mode (e.g. Control Patient, Cooling
or Rewarming).
1.
Verify Patient Temperature Accuracy
§
Confirm patient temperature using a secondary site.
§
Confirm that the primary temperature probe is properly
placed and registering an accurate temperature.
2.
Verify Patient / Water Temperature Control
§
Observe patient temperature.
Is the patient temperature increasing?
§
Observe water temperature.
Is the water temperature >31°C (87.8°F)?
Is the water temperature increasing?
§
If the water and/or patient temperatures are beginning to
increase, the system is working correctly. Continue to
monitor patient temperature.
3.
Verify Water Temperature Limits
§
In
Normothermia Settings
or
Hypothermia Settings
,
confirm that the water temperature high limit is set ≥36°C
(96.8°F).
§
If necessary, set water temperature high limit ≥36°C
(96.8°F).
§
Resume therapy.
§
Monitor water temperature and patient temperature.
§
If the water and/or patient temperatures are beginning to
increase, the system is working correctly. Continue to
monitor patient temperature.
4.
Verify Water Control – System Performance
§
Place unit in Manual mode, with water temperature set
> 38°C (100.4°F).
§
Monitor water temperature and patient temperature.
§
If the water temperature is not increasing to the
commanded temperature, the water heating system is not
working. Call BD Customer Support.
§
If the water temperature increases to the commanded
temperature, the water heating system is working.
5.
Verify Patient Control – System Performance
§
Place system back into the patient control mode.
§
Monitor water temperature and patient temperature.
§
If the water and/or patient temperatures are continuing to
increase, the system is working correctly. Continue to
monitor patient temperature.
§
If the water temperature and/or patient temperatures are not
increasing, call BD Customer Support.
11 Patient
Temperature 1 Below
Low Patient
Low Priority Alarm
In Normothermia Therapy: The
Patient Temperature 1 reading is
below the Low Patient Alarm
setting in Normothermia Settings.
In Hypothermia Therapy: The
Patient Temperature 1 reading is
below the Low Patient Alarm
setting in Hypothermia Settings.
If desired, adjust the
Low Patient Alarm
setting in
Normothermia
Settings
or
Hypothermia Settings.
12 Patient
Temperature 1 High
Low Priority Alarm
The Patient Temperature 1 reading
is above 39.5°C (103.1°F), and the
water temperature is above 39.5°C
(103.1°F) when the system is in
Manual Control mode.
Patient temperature is not
automatically controlled while in
Manual Control mode
Decrease Manual Mode water target temperature to a setting that is ≤
38.5°C (101.3°F).
See
VI. Operation Guide–Therapy Settings–Manual Control
for further
instructions.
13 Patient
Temperature 1 Low
Low Priority Alarm
The Patient Temperature 1 reading
is below 31°C (87.8°F), and the
water temperature is below 31°C
(87.8°F) when the system is in
Manual Control mode.
Patient temperature is not
automatically controlled while in
Manual Control mode
Increase Manual Mode water target temperature to a setting that is ≥ 32°C
(89.6°F).
See
VI. Operation Guide–Therapy Settings–Manual Control
for further
instructions.