Alarm
Problem
Solution
21 Patient
Temperature 2 High
Medium Priority
Alarm
The Patient Temperature 2 reading
is above 39.5°C (103.1°F), and the
water temperature is above 39.5°C
(103.1°F), and the system is
continuing to warm the patient
when the system is in a patient
control mode (e.g. Control Patient,
Cooling or Rewarming).
1.
Verify Patient Temperature Accuracy
§
Compare to Temp In 1 reading.
§
Confirm patient temperature using a secondary site.
§
Confirm that the secondary temperature probe is properly
placed and registering an accurate temperature.
2.
Verify Patient / Water Temperature Control
§
Observe patient temperature.
Is the patient temperature decreasing?
§
Observe water temperature.
Is the water temperature <39.5°C (103.1°F)?
Is the water temperature decreasing?
§
If the water and/or patient temperatures are beginning to
decrease, the system is working correctly. Continue to
monitor patient temperature.
3.
Verify Water Temperature Limits
§
In
Hypothermia Settings
or
Normothermia Settings
,
confirm that the water temperature low limit is set ≤10°C
(50°F).
§
If necessary, set water temperature low limit ≤10°C (50°F).
§
Resume therapy.
§
Monitor water temperature and patient temperature.
§
If the water and/or patient temperatures are beginning to
decrease, the system is working correctly. Continue to
monitor patient temperature.
4.
Verify Water Cooling – External Factors
§
Check environmental conditions that may affect cooling.
•
Are the machine vents blocked?
•
Is the room temperature too high?
•
Is the oxygen heated?
•
Are there hot lights or heating blankets on the
patient?
§
Remove any impediments.
§
Monitor water temperature and patient temperature.
§
If the water and/or patient temperatures are beginning to
decrease, the system is working correctly. Continue to
monitor patient temperature.
5.
Verify Water Control – System Performance
§
Place unit in Manual mode, with water temperature set
< 38°C (100.4°F).
§
Monitor water temperature and patient temperature.
§
If the water temperature is not decreasing to the
commanded temperature, the water cooling system is not
working. Call BD Customer Support.
§
If the water temperature decreases to the commanded
temperature, the water cooling system is working. Go to
Solution 6 to check the patient control performance.
6.
Verify Patient Control – System Performance
§
Place system back into the patient control mode.
§
Monitor water temperature and patient temperature.
§
If the water and/or patient temperatures are continuing to
decrease, the system is working correctly. Continue to
monitor patient temperature.
§
If the water temperature and/or patient temperatures are not
decreasing, call BD Customer Support.
22 Patient
Temperature 2 Above
High Patient
Low Priority Alarm
In Normothermia Therapy: The
Patient Temperature 2 reading is
above the High Patient Alarm
setting in Normothermia Settings.
In Hypothermia Therapy: The
Patient Temperature 2 reading is
above the High Patient Alarm
setting in Hypothermia Settings.
If desired, adjust the
High Patient Alarm
setting in
Normothermia
Settings
or
Hypothermia Settings.