Alarm
Problem
Solution
66 Non-Recoverable
System Error
Medium Priority
Alarm
Unable to disable pump power
(Control processor).
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
67 Non-Recoverable
System Error
Medium Priority
Alarm
Unable to disable pump power
(Monitor processor).
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
71 Non-Recoverable
System Error
Medium Priority
Alarm
Primary outlet water temperature
sensor out of range – high
resistance.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
72 Non-Recoverable
System Error
Medium Priority
Alarm
Primary outlet water temperature
sensor out of range – low
resistance.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
73 Non-Recoverable
System Error
Medium Priority
Alarm
Secondary outlet water
temperature sensor out of range –
high resistance.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
74 Non-Recoverable
System Error
Medium Priority
Alarm
Secondary outlet water
temperature sensor out of range –
low resistance.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
75 Non-Recoverable
System Error
Medium Priority
Alarm
Inlet water temperature sensor out
of range – high resistance.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
76 Non-Recoverable
System Error
Medium Priority
Alarm
Inlet water temperature sensor out
of range – low resistance.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
77 Non-Recoverable
System Error
Medium Priority
Alarm
Chiller water temperature sensor
out of range – high resistance.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
78 Non-Recoverable
System Error
Medium Priority
Alarm
Chiller water temperature sensor
out of range – low resistance.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
79 Non-Recoverable
System Error
Medium Priority
Alarm
Primary and secondary outlet
water temperature sensors differ
by greater than 1°C.
3.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
4.
If alarm persists, contact BD Customer Support.
80 Non-Recoverable
System Error
Medium Priority
Alarm
The control processor failed to
detect a simulated water
temperature fault.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
81 Non-Recoverable
System Error
Medium Priority
Alarm
The monitor processor failed to
detect a simulated water
temperature fault.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
82 Non-Recoverable
System Error
Medium Priority
Alarm
Processor fault.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
83 Non-Recoverable
System Error
Medium Priority
Alarm
Monitor processor communications
fault.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
84 Non-Recoverable
System Error
Medium Priority
Alarm
Control processor communications
fault.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.
86 Non-Recoverable
System Error
Medium Priority
Alarm
Power supply voltage fault.
1.
Turn Control Module Off. Wait 30 seconds and turn Control Module
On.
2.
If alarm persists, contact BD Customer Support.