Operator’s Manual
40
The Arctic Sun™ Temperature Management System unit is intended for use in
the electromagnetic environment specified below. The customer or the end
user of the Arctic Sun™ Temperature Management System should assure it is
used in such an environment.
Immunity
Test
IEC60601
test level
Compliance
Level
Electromagnetic
Environment – Guidance
Conducted
RF
IEC 61000-
4-6
Radiated RF
IEC 61000-
4-3
3 Vrms
6Vrms
(In ISM
Bands)
150 kHz to 80
MHz
80MHz to
2.7GHz
3Vrms
6Vrms
(In ISM Bands)
150 kHz to 80
MHz
3 V/m
80 MHz – 2,7
GHz
80 % AM at 1
kHz
Portable and mobile RF
communications
equipment should be
used no closer to any
part of the Arctic Sun™
Temperature
Management System
including cables, than the
recommended
separation distance
calculated from the
equation applicable to
the frequency of the
transmitter.
Recommended
separation distance
d = 1,2√P
d = 1,2√P
d = 2,3√P
where P is the maximum
output power rating of
the transmitter in watts
(W) according to the
transmitter manufacturer
and d is the
recommended
separation distance in
meters (m).
Field strengths from fixed
RF transmitter as
determined by an
electromagnetic site
survey, should be less
than the compliance level
in each frequency range.
Even if other equipment
complies with CISPR8
emission requirements,
interference may occur in
the vicinity of equipment
marked with the following
symbol:
1.3 EN/IEC 60601-1-2: Table 3:
Test
freq.
(MHz)
Band
(MHz)
Service
Modulation
Max
Power
(W)
Distance
(m)
Immunity
Test
Level
(V/m)
385
380 -
390
TETRA 400
Pulse
mod.
18 Hz
1.8
0.3
27
450
430 –
470
GMRS 460
FRS 460
FM +-5
kHz
deviation
1kHz
sine
2
0.3
28
710
704 -
787
LTE Band
13, 17
Pulse
mod.
217 Hz
0.2
0.3
9
745
780
810
800 -
960
GSM
800/900,
TETRA 800,
iDEN 820,
CMDA 850,
LTE band 5
Pulse
mod.
18 Hz
2
0.3
28
870
930
1720
1700 -
1990
GSM 1800,
CDMA 1900,
GSM 1900,
DECT,
LTE band 1,
3, 4, 25
UMts
Pulse
mod.
217 Hz
2
0.3
28
1845
1970
2450
2400 -
2570
Bluetooth,
WLAN,
802.11 b/g/n
RFID 2450
LTE band 7
Pulse
mod.
217 Hz
2
0.3
28
5240
5100 -
5800
WLAN
802.11 a/n
Pulse
mod.
217 Hz
0.2
0.3
9
5500
5785
1.4 EN/IEC 60601-1-2:2007 Sub-clause 5.2.2.2 Table 6:
Recommended separation distances between portable and mobile RF
communications equipment and the Arctic Sun™ Temperature Management
System unit
RF communications equipment can affect medical electrical equipment. The Arctic
Sun™ Temperature Management System unit is intended for use in an
electromagnetic environment in which radiated RF disturbances are controlled. The
customer or the user of the Arctic Sun™ Temperature Management System unit
can help prevent electromagnetic interference by maintaining a minimum distance
between the portable and mobile RF communications equipment (transmitters) and
the Arctic Sun™ Temperature Management System unit as recommended below,
according to the maximum output power of the communications equipment.
Rated maximum
output power of
transmitter in
watts (W)
Separation distance according to frequency of transmitter in
meters (m)
150kHz to 80MHz
d = 1.2√P
80MHz to
800MHz
d = 1.2√P
800MHz to
2.5GHz
d = 2.3√P
0.01
0.12
0.12
0.23
0.1
0.38
0.38
0.73
1.0
1.2
1.2
2.3
10
3.8
3.8
7.3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be estimated using the
equation applicable to the frequency of the transmitter, where P is the maximum
output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects and
people.