T r o u b l e s h o o t i n g
Power Module DN
Version 09.2019
Page 34/43
9.8 Demagnetization pulse check
•
Prepare suitable current clamp
•
Open power module
•
Clamp current clamp around power cable a
Coil module connector
•
Trigger pulse and record current over entire pulse length
•
Maximal current (RMS) corresponds to the value in the inspection protocol
•
Take a picture of the pulse (example see below) and send incl. setup current clamp
(mV/A) to Maurer Magnetic AG for analysis
Figure 9: Measurement demagnetization pulse
9.9 Malfunction external coil fan at Fan connector
Inspect fuse 1FT2: Fuse
Fan connector