I n s t a l l a t i o n
Power Module DN
Version 09.2019
Page 13/43
4
Installation
Mount device on and flat ground, stable and flat surface.
•
Ensure safe operation of device at all times
•
Operate power module only upright.
Do not operate in horizontal position
.
Connection and disconnection of the coil module only with main switch off.
Installation and commissioning of the device only by skilled personell.
•
Connect only to power source with correctly installed protection conductor
•
Connect with regard to local regulations
Improper connection can damage the machine.
•
Installation with off-state power source
•
Installation with main switch off
Interface signals maximal current load 100mA
D-Sub15 PIN14/15: 0...10 VDC Output maximal current load 1 mA
Do not connect D-Sub interface with computer interface verbinden. Pc and power
module may be damaged.
4.1 Transport
Do not lift components of more than 25 kg by hand.
Do not lift coils mounted in shielding chambers on the shielding chamber.
•
Lift with support from below
Power module can be lifted and transported by two persons.
4.2 Installation process
•
Mounting of coil
•
Mounting of power module
•
Check fastening torque of power connections
•
Connection power supply power module
•
Connection
Coil module connector
and
Coil module monitoring M8
•
Connection
Fan connector
(if applicable; MM VE)
•
Optional: Connection
D-Sub15 Control Interface
•
Optional: Connection
Power selection interface D-Sub9
(DN750-1850)
•
Optional: Connection photo sensor on
Sensor connector M12