Rad-97
Chapter 4: Operation
www.masimo.com
77
Masimo
Options
Description
Factory Default
Settings
User Configurable
Settings
Offset
Amount
See
In Vivo Adjustment
Overview
on page 61.
0.0 when turned
on
-9.0% to +9.0% in steps
of 0.1%
ORi Settings
ORi is currently not available in the U.S.A. and territories relying on FDA market clearance.
From the
ORi Settings
screen, access the following screens:
ORi Alarms
on page 77
Trends
on page 97
About Parameter Information
on page 78.
ORi Alarms
From the
ORi Alarms
screen, access the following screens:
Options
Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
Low Limit
Low Limit is the lower
threshold that triggers an
alarm.
Medium Off
Off, or 0.01 to 0.98
in steps of 0.01
Trending
Down Alarm
Trending Down Alarm is
displayed when a rapid
decrease in ORi is measured.
N/A
Off
On or Off
Silence
Duration
Sets the amount of time that
the alarm is silenced.
N/A
2 minutes
30 seconds or 1, 2,
or 5 minutes