Safety Information, Warnings and Cautions
www.masimo.com
19
Masimo
MRI electrically powered wheelchair, etc. When used in the presence of potential RFI devices,
consideration should be taken to maximize separation distances and to observe for any
potential signs of interference such as loss of communication or reduced Wi-Fi signal
strength.
CAUTION:
To ensure that alarm limits are appropriate for the patient being monitored, check
the limits each time Rad-97 is used.
CAUTION:
Replace the cable or sensor when a replace sensor or when a low SIQ message is
consistently displayed while monitoring consecutive patients after completing the low SIQ
troubleshooting steps listed in the troubleshooting section.
Note:
Cables and sensors are provided with X-Cal™ technology to minimize the risk of
inaccurate readings and unanticipated loss of patient monitoring. Refer to the Cable or
Sensor DFU for the specified duration of patient monitoring time.
Note:
SpHb readings may be inaccurate for patients with conditions that may cause edema
at the measurement site (eg. kidney disease, pregnancy, etc.).
Note:
Physiological conditions that result in loss of pulsatile signal may result in no SpO
2
,
SpHb, SpOC, SpCO, SpMet, RRp, and ORi readings.
Note:
Rad-97 is provided with a Wi-Fi signal indicator as an indication of Wi-Fi
communication.
Note:
Rad-97’s alarm capabilities have been designed to be independent of the Wi-Fi
communication feature in order to preserve Rad-97's primary alarms.
Note:
Always charge Rad-97 when it is not in use to ensure that the Rad-97 Battery Module
remains fully charged.
Note:
All batteries lose capacity with age, thus the amount of run time at Low Battery will
vary depending upon the age of the Battery Module.
Note:
A functional tester cannot be used to assess the accuracy of Rad-97.
Note:
When monitoring acoustic respiration, Masimo recommends minimally monitoring
both oxygenation (SpO
2
) and respiration (RRa).
Note
: When using the Maximum Sensitivity setting, performance of the "Sensor Off"
detection may be compromised. If the Rad-97 is in this setting and the sensor becomes
dislodged from the patient, the potential for false readings may occur due to environmental
"noise" such as light, vibration, and excessive air movement.
Noninvasive Blood Pressure
WARNING:
Before applying the cuff on the patient, confirm the cuff size is appropriate.
WARNING:
When a blood pressure measurement error code occurs, any blood pressure values
reported should be disregarded.
CAUTION:
If the blood pressure cuff is on the same limb as monitoring equipment (i.e., pulse
oximeter probe), the pressurization within the cuff can cause temporary loss of function of the
monitoring equipment.
CAUTION:
Blood pressure measurements can be affected by the patient's position,
physiological condition, and environmental factors.
Содержание Rad-97
Страница 1: ...Operator s Manual Rad 97 Pulse CO Oximeter...
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 36: ......
Страница 104: ......
Страница 110: ......
Страница 118: ......
Страница 124: ......
Страница 146: ......
Страница 168: ......
Страница 178: ......
Страница 186: ......
Страница 187: ......
Страница 188: ...www masimo com 38281 LAB 9275B 0917...