Rad-97
Chapter 11: Specifications
www.masimo.com
159
Masimo
Guidance and Manufacturer's Declaration - Electromagnetic Immunity
IEC 61000-4-2
+8 kV air
humidity should be at least 30%.
Electrical fast
transient/ burst
IEC 61000-4-4
+/- 2 kV
for power
lines
+/- 1 kV
for input/
output
lines
+/- 2 kV for
power lines
+/- 1 kV for
input/ output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
+/-1 kV
line(s) to
line(s)
+/- 2 kV
line(s) to
earth
+/-1 kV line(s)
to line(s)
+/- 2 kV line(s)
to earth
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions, and
voltage variations
on power supply
input lines
100% dip
in mains
voltage for
0.5 cycle
60% dip
in mains
voltage for
5 cycle
30% dip
in mains
voltage for
25 cycle
100% dip in
mains voltage
for 0.5 cycle
60% dip in
mains voltage
for 5 cycle
30% dip in
mains voltage
for 25 cycle
Mains power quality should be that of a
typical commercial or hospital
environment.
Power frequency (50
/ 60 Hz) magnetic
field.
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields
should be at levels characteristic of
typical location in a typical hospital
environment.
Portable and mobile RF
communications equipment should be
used no closer to any part of the ME
Equipment, including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation distance
Содержание Rad-97
Страница 1: ...Operator s Manual Rad 97 Pulse CO Oximeter...
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 36: ......
Страница 104: ......
Страница 110: ......
Страница 118: ......
Страница 124: ......
Страница 146: ......
Страница 168: ......
Страница 178: ......
Страница 186: ......
Страница 187: ......
Страница 188: ...www masimo com 38281 LAB 9275B 0917...