Rad-97
Chapter 4: Operation
www.masimo.com
74
Masimo
SpOC Settings
From the
SpOC Settings
screen, access the following screens:
SpOC Alarms
on page 74
Trends
on page 97
About Parameter Information
on page 78
SpOC Alarms
From the
SpOC Alarms
screen, access the following screens:
Options
Description
Alarm
Priority
Factory
Default
Settings
User Configurable
Settings
High Limit High Limit is the upper
threshold that triggers an
alarm.
Medium
25
2 ml/dl to 34 ml/dl in
steps of 1 ml/dl, or Off
Low Limit Low Limit is the lower
threshold that triggers an
alarm.
High
10
Off, or 1 ml/dl to 33
ml/dl in steps of 1
ml/dl
Silence
Duration
Sets the amount of time
that the alarm is silenced.
NA
2 minutes
30 seconds or 1, 2, or 5
minutes
SpMet Settings
From the
SpMet Settings
screen, access the following screens:
SpMet Alarms
on page 75
In Vivo for SpMet
on page 75
Trends
on page 97
About Parameter Information
on page 78