Rad-97
Safety Information, Warnings and Cautions
www.masimo.com
16
Masimo
WARNING:
SpO
2
is empirically calibrated in healthy adult volunteers with normal levels of
carboxyhemoglobin (COHb) and methemoglobin (MetHb).
WARNING:
Optical, pleth-based measurements (e.g. SpO
2
, SpHb, SpOC, SpMet, SpCO, RRp,
and ORi) can be affected by the following:
•
Improper sensor application or use of use of incorrect sensor.
•
Blood pressure cuff applied to the same arm as the sensor site.
•
Intravascular dyes such as indocyanine green or methylene blue.
•
Venous congestion.
•
Abnormal venous pulsations (e.g. tricuspid value regurgitation, Trendelenburg
position).
•
Abnormal pulse rhythms due to physiological conditions or induced through
external factors (e.g. cardiac arrhythmias, intra-aortic balloon, etc.).
•
Externally applied coloring and texture such as nail polish, acrylic nails, glitter, etc.
•
Moisture, birthmarks, skin discoloration, nail aberration, deformed fingers, or
foreign objects in the light path.
•
Elevated levels of bilirubin.
•
Physiological conditions that can significantly shift the oxygen disassociation
curve.
•
A physiological condition that may effect vasomotor tone or changes in vasomotor
tone.
WARNING:
Inaccurate SpO
2
readings may be caused by:
•
Elevated levels of COHb and/or MetHb.
•
Severe anemia.
•
Extremely low arterial perfusion.
•
Excessive induced motion.
•
Hemoglobinopathies (qualitative defects including sickle cell) and Hemoglobin
synthesis disorders (Quantitative defects such as Thalassemias).
WARNING:
Inaccurate SpHb and SpOC readings may be caused by:
•
Low arterial perfusion.
•
Motion induced artifact.
•
Low arterial oxygen saturation levels.
•
Elevated COHb and/or MetHb levels.
•
Hemoglobinopathies (qualitative defects including sickle cell) and Hemoglobin
synthesis disorders (quantitative defects such as Thalassemias).
•
Severe anemia.
WARNING:
Inaccurate SpCO readings may be caused by:
•
Elevated methemoglobin levels in the range of >15%.
Содержание Rad-97
Страница 1: ...Operator s Manual Rad 97 Pulse CO Oximeter...
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 36: ......
Страница 104: ......
Страница 110: ......
Страница 118: ......
Страница 124: ......
Страница 146: ......
Страница 168: ......
Страница 178: ......
Страница 186: ......
Страница 187: ......
Страница 188: ...www masimo com 38281 LAB 9275B 0917...