Rad-97
Safety Information, Warnings and Cautions
www.masimo.com
18
Masimo
•
Hypotension, severe vasoconstriction, severe anemia, or hypothermia.
WARNING:
ORi is not intended as a replacement for SpO
2
monitoring, PaO
2
monitoring, or as
a sole indicator of the patient condition.
WARNING:
Wireless communication of alarms to a secondary monitoring station should not
be relied upon as a primary alarm.
CAUTION:
ORi may not indicate additional changes in oxygen states above 200 mmHg of
PaO
2
.
CAUTION:
If using Rad-97 during full body irradiation, keep the sensor out of the radiation
field. If the sensor is exposed to the radiation, the reading might be inaccurate or the device
might read zero for the duration of the active irradiation period.
CAUTION:
When patients are undergoing photodynamic therapy they may be sensitive to
light sources. Pulse oximetry may be used only under careful clinical supervision for short
time periods to minimize interference with photodynamic therapy.
CAUTION:
High ambient light sources such as surgical lights (especially those with a xenon
light source), bilirubin lamps, fluorescent lights, infrared heating lamps, and direct sunlight
can interfere with the performance of the sensor.
CAUTION:
To prevent interference from ambient light, ensure that the sensor is properly
applied, and cover the sensor site with opaque material, if required. Failure to take this
precaution in high ambient light conditions may result in inaccurate measurements.
CAUTION:
If the Low Perfusion message is frequently displayed, find a better perfused
monitoring site. In the interim, assess the patient and, if indicated, verify oxygenation status
through other means.
CAUTION:
To minimize radio interference, other electrical equipment that emits radio
frequency transmissions should not be in close proximity to Rad-97.
CAUTION
:
Do not place the Rad-97 near electrical equipment that may affect the device,
preventing it from working properly.
CAUTION:
Failure to charge Rad-97 promptly after a Low Battery alarm may result in the
device shutting down.
CAUTION:
Do not connect to an electrical outlet controlled by a wall switch or dimmer.
CAUTION:
In order to establish and maintain Rad-97’s minimum Quality of Service, the
following network specifications should be met before and after installation:
•
Wired Network Connection
During Ping Test, passing result if:
a. At least 98% of packets have latency ≤ 30 milliseconds, and
b. No more than 2 % packets loss.
•
Wireless Network Connection
During Ping Test, passing result if:
a. At least 98% of packets have latency ≤ 100 milliseconds,
b. No more than 2 % packets loss, and
c. Primary access point signal strength at least -67 dBm.
CAUTION:
The wireless quality of services may be influenced by the presence of other devices
that may create radio frequency interference (RFI). Some RFI devices to consider are as
follows: electrocautery equipment, cellular telephones, wireless PC and tablets, pagers, RFID,
Содержание Rad-97
Страница 1: ...Operator s Manual Rad 97 Pulse CO Oximeter...
Страница 2: ......
Страница 8: ......
Страница 10: ......
Страница 14: ......
Страница 36: ......
Страница 104: ......
Страница 110: ......
Страница 118: ......
Страница 124: ......
Страница 146: ......
Страница 168: ......
Страница 178: ......
Страница 186: ......
Страница 187: ......
Страница 188: ...www masimo com 38281 LAB 9275B 0917...