Rad-97
Chapter 11: Specifications
www.masimo.com
150
Masimo
Parameter Resolution
SpMet
0.1%
SpOC
1.0 ml/dL
RRp**
1 RPM
* Breaths per minute is denoted as rpm on device display.
** Parameter currently not available in the U.S.A. and territories relying on FDA market
clearance.
Noninvasive Blood Pressure (NIBP) Specifications
Measurement Range
Measurement Patient Population Display Range
Systolic
Adult
40-260 mmHg
Pediatric
40-230 mmHg
Neonatal
40-130 mmHg
Diastolic
Adult
20-200 mmHg
Pediatric
20-160 mmHg
Neonatal
20-100 mmHg
MAP
Adult
26-220 mmHg
Pediatric
26-183 mmHg
Neonatal
26-110 mmHg
Pulse Rate (PR) All
30-220 bpm