Press the arrow keys to change the blinking value.
Press the GLP/ACCEPT key to accept the value or the CAL key to escape.
Press the arrow keys to select the next/previous parameter.
Press the SETUP key to exit SETUP menu at any time.
The following table lists the SETUP parameters, their valid values range and the factory settings
(default):
Notes
Notes
Notes
Notes
Notes:
•If “dISP” option is ON, the tags corresponding to the calibrated buffers are displayed on the
LCD while in pH measurement mode.
•To select the right ion charge, different ion types and their charge are listed in the table
below:
In Id
Instrument ID
0000 to 9999
0000
bEEP
Beep Status
ON/OFF
OFF
OFF dAY
Alarm Time Out
OFF or 1 to 14 days
OFF
TIME
Time (hh:mm)
00:00 to 23:59
00:00
Item
Description
Valid values
Default
dISP
Display Cal Buffers
ON/OFF
ON
IonCG
Ion Charge
UndF or -2; -1; 1; 2
UndF
tEMP
Temperature Unit
°
C or
°
F
°
C
DATE
Date (MM.DD.YYYY)
01.01.2000-12.31.2099
01.01.2005
bAud
Baud Rate
600;1200;2400;4800;9600
2400
ION CHARGE
ION types
-2 (divalent anions)
S, CO
3
-1 (monovalent anoins)
F, Cl, Br, I, CN, SCN, ClO
4
, NO
3
1 (monovalent cations)
H, Na, K, Ag, NH
4
2 (divalent cations)
Mg, Ca, Ba, Cd, Cu, Pb
UndF
Undefined ion