• When the reading is stable and close to the selected standard, the “READY” and
“ACCEPT” tags will blink.
• Press the GLP/ACCEPT key to confirm calibration.
• The calibrated value will be displayed on the primary LCD and the second expected
standard value on the secondary LCD.
Note
Note
Note
Note
Note: The instrument will automatically skip the standard used for the first point.
• After the first calibration point is accepted, immerse the ISE electrode approximately 4 cm
(1½”) into the second calibration solution.
• If necessary, press the UP and DOWN arrow keys to select the standard value.
• The “WAIT” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected standard, the “READY” and “ACCEPT”
tags will blink.
• Press the GLP/ACCEPT key to confirm calibration.
• The instrument stores the calibration value and returns to normal measurement mode.
Notes
Notes
Notes
Notes
Notes:
•If the mV value is out of the mV range (±2000), the closest full-scale value will be displayed
blinking.