GENERAL DESCRIPTION
Thank you for choosing Martini Instruments. This instruction manual will provide you the
necessary information for correct use of the meter.
Mi 160
Mi 160
Mi 160
Mi 160
Mi 160 is a logging microprocessor-based pH, ORP, ISE and temperature bench meter.
This meter is provided with a series of new diagnostic features which add an entirely
new dimension to the measurement of pH/ORP/ISE, by allowing the user to dramati-
cally improve the reliability of the measurement:
• 7 memorized buffers (pH 1.68, 4.01, 6.86, 7.01, 9.18, 10.01 and 12.45) for pH
calibration
• pH calibration up to 3 points
• ISE calibration up to 2 points; six standard solutions available: 0.01, 0.1, 1, 10,
100, 1000 ppm
• Messages on the LCD to make the calibration easy and accurate
• Relative mV feature
• GLP feature, to view last calibration data for pH or ISE
• User-selectable alarm time out to alert the user that too much time elapsed since
the last pH calibration
• Log-on-demand (50 samples for each range: pH, mV/Rel mV, ISE)
Moreover, it offers an extended temperature range from –20 °C (–4 °F) to 120 °C
(248 °F), using MA 831R
MA 831R
MA 831R
MA 831R
MA 831R interchangeable temperature probe.
For accurate measurements, use
the
electrode holder supplied with the meter.
This Bench Meter is supplied with:
• MA 917B/1 pH Electrode
• MA 831R Temperature Probe
• MA 9315 Electrode Holder
• M 10004 pH 4.01 Sachet Buffer Solution
• M 10007 pH 7.01 Sachet Buffer Solution
• M 10010 pH 10.01 Sachet Buffer Solution
• M 10016 Sachet Electrode Cleaning Solution
• Mi 5200 Application Software
• MA 9350 RS232 Connector cable (2 meters)
• Graduate Pipet •12 VDC Adapter • Instruction Manual