RELATIVE mV CALIBRATION
• Press the CAL key when the instrument is in mV / Rel mV measurement mode. The
“CALIBRATION” tag will appear on the LCD.
• Relative mV value is displayed on the primary LCD and the absolute mV value on the
secondary LCD.
• The “WAIT” tag will blink until the reading is stable.
• When the absolute reading is stable and in measurement range, the “READY” and
“ACCEPT” tags blink on the LCD, asking for confirmation.
• If the reading is out of range, the absolute mV value and the “WRONG” tag will blink.
• Press the GLP/ACCEPT key to confirm the calibration. The instrument enters Rel mV
measurement mode.
Notes
Notes
Notes
Notes
Notes:
•If a Rel mV calibration is performed, the range changes from mV to Rel mV.
•To return to mV measurement mode, clear the Rel mV calibration by pressing LOG/CLR
after entering calibration mode. The “CLr CAL” message will appear on the LCD for one
second and the instrument will enter mV measurement mode.