• If “no BUFFER” message appears on the LCD, the instrument informs you that
calibration was performed in only one point.
• If a one-point calibration is performed after a two-point calibration, the instrument
will keep the old slope.
• Press the GLP/ACCEPT key at any moment and the instrument will return to measure-
ment mode.
• If calibration has not been performed for the selected range, the instrument displays
“no CAL” message blinking.
LOGGING
Up to 50 readings can be stored into memory for each measurement range (pH, mV/Rel mv
and ISE).
LOGGING THE CURRENT DATA
LOGGING THE CURRENT DATA
LOGGING THE CURRENT DATA
LOGGING THE CURRENT DATA
LOGGING THE CURRENT DATA
To store the current reading into memory press the LOG/CLR key while in measurement
mode.
The instrument will display the current date (MM.DD) on the primary LCD and the record
number on the secondary LCD, along with “LOG” tag (see example below: record No. 12,
dated May 18).