Instruction Manual Mi 160 Bench Meter
10
10
10
10
10
• If necessary, press the UP and DOWN arrow keys to select a different buffer value.
• The “WAIT” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, the “READY” and “ACCEPT” tags
will blink on the LCD.
• Press the GLP/ACCEPT key to confirm calibration.
• The calibrated value will be displayed on the primary LCD and the second expected buffer
value on the secondary LCD, along with the tag of the buffer already calibrated.
• After the first calibration point is accepted, immerse the pH electrode and the temperature
probe approximately 4 cm (1½”) into the second buffer solution and stir gently. The
temperature probe should be close to the pH electrode.
• If necessary, press the UP and DOWN arrow keys to select a different buffer value.
• The “WAIT” tag will blink on the LCD until the reading is stable.
• When the reading is stable and close to the selected buffer, the “READY” and “ACCEPT” tags
will blink on the LCD.
• Press the GLP/ACCEPT key to confirm calibration.
• The calibrated value will be displayed on the primary LCD and the third expected buffer value
on the secondary LCD, along with the tags of the buffers already calibrated.