My meter "
Total
" isn't right
Symptom
Cause
Remedy
My meter total "creeps" up
Normally occurs due to moving
water - leaky gate, foot valve
Check seals of gates and
valves. Consider installing a
second gate to isolate the sys-
tem.
Entrapped aquatic life in pipe
Adjust
"Filters"
using
Water "sloshing" in pipe around
sensor level
Make sure water level recedes
fully, or pipe stays completely
full. Consider installing a
second gate to isolate the sys-
tem.
Electromagnetic interference
(EMI) causing false "flow"
Find source of EMI and rectify
if possible
My total is not increasing
Total channel set up as "
Neg-
ative only
"
Check "
" in
My total is not decreasing
Total channel set up as "
Pos-
itive only
"
Check "
" in
I can't reset the total
Total channel has been set up
as "
Non-resettable
"
Check "
" in
My total "rolls over" too
quickly
Total units are too small for
application
Change
to larger
units
My total is wrong!
Total units are not what's expec-
ted
in
Total has "rolled over" without
me realising it!
MACE XCi User Manual - 251 -
Содержание HydroMace XCi
Страница 1: ......
Страница 13: ...MACE XCi User Manual 13...
Страница 19: ...MACE XCi User Manual 19...
Страница 33: ...MACE XCi User Manual 33...
Страница 36: ...MACE XCi User Manual 36...
Страница 167: ...NOTE None of these parameters can be edited MACE XCi User Manual 167...
Страница 182: ...MACE XCi User Manual 182...
Страница 187: ...MACE XCi User Manual 187...
Страница 200: ...NOTE In the example the Net flowrate will be the sum of Flow Rate 1 Flow Rate 2 Flow Rate 3 MACE XCi User Manual 200...
Страница 260: ...5 Tighten all three screws 6 Reconnect the battery cable assembly to the backplane board MACE XCi User Manual 260...
Страница 265: ...MACE XCi User Manual 265...