Units text
This sets the units on the logged value. This is a text field and is user defined.
Log pulse event
If this check box is ticked a “
Status
” message is logged in the data file when the event occurs.
Factor
This field scales the pulse to match the units being used. For example, if 1 kilolitre pulses and megalitre units
are used, the factor would be 0.001 where as if 1 kilolitre pulses and kilolitre units are used, the factor would
be 1.0
Alternatively when the input is from a rain gauge, if the bucket size is 0.25mm then each pulse will cause the
totalizer to increment by that amount
(Eg. 0.25mm, 0.50mm, 0.75mm.....)