New scaled value
The current flow rate value in the configured “
Flowrate units
” is entered here.
2. With the sensor connected to the Pulse card, a steady stream flow in the pipe should be obtained.
Over a number of minutes, the “
Current pulse rate
” should stabilise.
3. Once the “
Current pulse rate
” has stabilised, enter the actual flow rate value in the “
Flowrate units
”
configured into the “
New scaled value
” field. Click “
Accept
” and the XCi device will calculate the
new “
K-factor
” to apply to incoming pulses.
MACE XCi User Manual - 227 -