l
Version of firmware installed in the XCi device;
l
Time and date of the clock in the XCi device;
l
Battery status and charge (voltage) of the XCi device;
l
Number of channels being recorded;
l
The names of each channel being recorded;
l
The units of measure used for each channel being recorded;
l
‘Points’ refers to the number of data lines in the downloaded file;
l
‘Interval’ refers to the time in seconds between each time stamp (logging interval).
Status messages
Status messages are defined by the use of an asterisk (*) immediately following a time stamp. For example,
in the download image above the following message appears which tells the user the date and time the unit
was started.
2011/08/24 16:28:52, *Note, "Unit started"
Table of Status messages
Status message
Description
Unit Started
Device was started
Unit stopped
Device was stopped
No external power
No external power was detected. If a solar panel is used this will
occur at sunset
External power restored
External power is detected. If a solar panel is used this will occur at
sunrise.
Low Battery
Time at which battery voltage fell below 11.5V
Battery normal
Time at which battery voltage exceeded 11.5V
Flat battery condition
Battery voltage at download time is less than 10V
Battery was flat
Battery was less than 10V
Sensor error
Any configured module not detected
Device reset #0*
Power up reset occurred. Generally occurs when the battery has
been changed without first "
Stopping
" the unit
Device reset #1*
Clock reset. Generally occurs if a command line error occurs when
talking to the Real-time clock (RTC).
MACE XCi User Manual - 126 -
Содержание HydroMace XCi
Страница 1: ......
Страница 13: ...MACE XCi User Manual 13...
Страница 19: ...MACE XCi User Manual 19...
Страница 33: ...MACE XCi User Manual 33...
Страница 36: ...MACE XCi User Manual 36...
Страница 167: ...NOTE None of these parameters can be edited MACE XCi User Manual 167...
Страница 182: ...MACE XCi User Manual 182...
Страница 187: ...MACE XCi User Manual 187...
Страница 200: ...NOTE In the example the Net flowrate will be the sum of Flow Rate 1 Flow Rate 2 Flow Rate 3 MACE XCi User Manual 200...
Страница 260: ...5 Tighten all three screws 6 Reconnect the battery cable assembly to the backplane board MACE XCi User Manual 260...
Страница 265: ...MACE XCi User Manual 265...