Maintenance
42
© Lutz-Jesco GmbH 2020
BA-42840-02-V03
Water sampling station
EASYPRO COMPACT
Operating instructions
Resources required:
@
Slotted screwdriver
@
New fuse: 5 x 20 mm, 3.15 A, 250 V (delay)
Perform the following work steps:
1.
The fuse holder in the form of a bayonet catch is located at the
bottom right-hand side, above the clamps for PE, N and L with the
marking “Fuse”. Use the slotted screwdriver to press the catch down-
wards and then turn it leftwards.
2.
Remove the fuse.
3.
Replace the fuse and fix it in place by turning the catch clockwise.
ü
Fuse has been replaced.
11.9 Resetting the settings
i
The instructions differentiate between the internal factory
settings and the device configuration.
The factory settings contain the basic configuration of the
device hardware and cannot be changed.
The configuration file (*.SET) contains the individual device
configuration. You can change, save and load the individual
settings.
Reset to the factory settings
You can now reset the device to its factory settings. This deletes the
configuration. You must then either load a configuration file or perform the
configuration manually.
Perform the following work steps:
1.
Working in the main menu, navigate to System > Service > Device.
2.
Press “factory settings”.
3.
Confirm with “Yes”.
4
The configuration will be deleted. You must proceed with the
following section.
ü
All factory default settings will be reset.
Reset the configuration
The device configuration will be saved in *.SET files. A factory-set
configuration file with standard settings is already present. You can
change these or save your personal configuration in new files.
Recommendation:
Leave the factory-set configuration file unchanged
and save your personal configuration in a new file. Given problems with
the configuration, this enables you to return to a functioning configuration
quickly.
Perform the following work steps:
1.
Working in the main menu, navigate to System > Settings >
Configuration.
2.
Select an existing configuration file.
3.
Click “Load” to confirm.
4
The device configuration returns to the saved state.
ü
Load the old configuration.
11.10 Finishing maintenance
Perform the following work steps:
1.
Make a note of the date and scope of the maintenance performed.
2.
In the “Service” menu, navigate to the “Service entry” tab. Enter your
company name and notes about the maintenance. Activate the
reminder function and enter a date for the next service. Confirm with
the “Save” button .
4
Your service action has been saved in the logfiles.
3.
To restart the system, proceed in accordance with the instructions in
section 8 “Commissioning of the water sampling station” on
page 23.
ü
Maintenance completed.