Shutdown
39
© Lutz-Jesco GmbH 2020
Subject to technical changes.
200116
BA-42840-02-V03
Water sampling station
EASYPRO COMPACT
Operating instructions
10 Shutdown
10.1 Short-term shutdown
Short-term shutdown involves deactivation for the calibration of the
measured values or for maintenance work, for example. For long-term
shutdown, see section 10.2 “Long-term shutdown” on page 39.
Perform the following work steps:
1.
Close the inflow tap (Fig. 1 “Water sampling station EASYPRO
COMPACT” on page 8, pos. 8) of the water sampling station or
switch off the sample water supply externally. This interrupts the
measurement and the control of the
TOPAX
®
MC
controller, and the
sample water circuit can be opened.
2.
Disconnect the
TOPAX
®
MC
controller's external power supply.
3.
Perform the planned work on the water sampling station.
4.
Reconnect the
TOPAX
®
MC
controller's external power supply.
5.
Open the inflow tap of the water sampling station or switch the
sample water supply back on externally. It starts the measurement
and regulation.
ü
The EASYPRO COMPACT water sampling station has been
successfully shutdown temporarily and put back into operation
again.
10.2 Long-term shutdown
This section describes the required actions in the event of long-term shut-
down in order to ensure the functionality of the water sampling station.
Perform the following work steps:
1.
Close the inflow tap (Fig. 1 “Water sampling station EASYPRO
COMPACT” on page 8, pos. 8) of the water sampling station or
switch off the sample water supply externally. This interrupts the
measurement and the control of the
TOPAX
®
MC
controller, and the
sample water circuit can be opened.
2.
Disconnect the
TOPAX
®
MC
controller's external power supply.
3.
Pinch off all cables in the
TOPAX
®
MC
controller.
4.
Uncouple the ingoing hose connections from the inflow tap and from
the sample water filter (Fig. 1 “Water sampling station EASYPRO
COMPACT” on page 8, pos. 9), as well as the outgoing hose
connection from the outflow tap (pos. 4).
5.
Unscrew the temperature sensor (pos. 5) from the station, open the
extraction point and bleed the water from the water sampling station.
6.
Now immediately take the measuring cells out of the valves and store
them stood upright in a 3-molar potassium chloride solution.
ü
The EASYPRO COMPACT water sampling station has success-
fully been shutdown for the long term.
10.3 Storage
Pre-conditions for actions:
ü
The device has been shut down in accordance with the section 10.2
„Long-term shutdown“.
Storing the water sampling station correctly extends its service life. You
should avoid negative influences such as extreme temperatures, high
humidity, dust, chemicals, etc.
Unscrew the Water sampling station from the wall holder and store them
away.
Ensure ideal storage conditions where possible:
n
The storage place must be cold, dry, dust-free and generously
ventilated
n
Temperatures b 0 °C and + 50 °C
n
Relative air humidity must not exceed 90 %
10.4 Transportation
Pre-conditions for actions:
ü
The water sampling station has been shut down in accordance with
chapter 10.2 „Long-term shutdown“.
n
The water sampling station may only be transported in an empty state.
n
Avoid heavy blows at all costs.
If the device is sent back to the manufacturer, please follow sections 15
“EU Declaration of Conformity” on page 52 and section 17 “Warranty
claim” on page 54.
10.5 Disposal
n
The device must be disposed of in accordance with applicable local
laws and regulations. It should not be disposed of as domestic waste!
n
As the disposal regulations may differ from country to country, please
consult your supplier.
n
In Germany, the manufacturer must provide free-of-charge disposal
provided the device has been safely returned along with a declaration
of no objection (see chapter 16 “Declaration of no objection” on
page 53).