30 - Preparation for Testing
3.4 Sample application using other sample types
Refer to LumiraDx Test Strip Product Inserts for information on available sample types per
test, and procedures to transfer other sample types to the Sample Application Area of
the Test Strip. Use the recommended transfer procedure to dispense a small sample of
one drop onto the circular Sample Application Area on top of the inserted Test Strip.
3.5 Quality Control solution application
Refer to LumiraDx Quality Control Pack Inserts for information on testing with LumiraDx
Quality Controls.
Do NOT
apply more than one drop of sample.
Do NOT
touch the Test Strip until the test has finished and the result is displayed.
Applying a transfer tube sample to the
Test Strip
Hold the filled transfer tube vertically over
the Sample Application Area and gently
squeeze the bulb or depress the plunger
to move the sample to the tip of the
transfer tube.
Allow the blood drop to touch the sample
application area without releasing the
pressure on the bulb to apply the sample.
When the sample is detected, the
Instrument will sound, and a confirmation
message will be displayed.
Dispose of the lancet in the appropriate
clinical waste. Clean the patient’s finger
with a clean tissue and apply slight
pressure.
If you need to retest, use a new Test Strip
and lancet, and a different finger.
Hold the transfer tube vertically and gently
squeeze the bulb to apply the blood drop
to the Sample Application Area.
Hold the transfer tube vertically and gently
depress the plunger to apply the blood
drop to the Sample Application Area.
Содержание SARS-CoV-2
Страница 31: ......
Страница 32: ...Platform User Manual ...
Страница 33: ......
Страница 41: ......
Страница 96: ...Instrument Operation 63 A test which fails the OBC will appear in the Invalid filter of the Results History ...
Страница 109: ...76 Ancillary Devices LumiraDx Desktop Stand ...
Страница 117: ...84 ...
Страница 118: ......