4.7 Quality Control
The LumiraDx Instrument has a number of built-in QC functions. In addition to
these, LumiraDx Quality Controls can be used to meet compliance requirements
as required.
To perform QC testing with LumiraDx Quality Controls, you will need:
•
LumiraDx Instrument
•
LumiraDx Test Strips
•
LumiraDx Quality Controls
Preparing for a QC test requires the same steps as preparing to perform a
patient test. The only difference in the test method is the use of Quality Controls
instead of patient samples.
Instruments being managed in a Connected mode may have the Quality
Control policy set to ‘Mandatory’ by in the Connect Manager settings. This
means that users and administrators are not able to perform a patient test when
a Quality Control test for that particular test is due. The timing and frequency of
mandatory Quality Control tests are configurable in Connect Manager.
48 - Instrument Operation
14. Remove and dispose of Test Strip
The Instrument will prompt to remove the Test
Strip.
Gently pull the blue label end of the Test Strip to
remove and dispose.
Disinfect the Instrument between patient
tests. Refer to the “Cleaning and Disinfecting”
chapter in this Platform User Manual for further
information.
Close the Instrument door.
Dispose of all Test Strips used for
patient testing safely in accordance
with local regulations and procedures.
Содержание SARS-CoV-2
Страница 31: ......
Страница 32: ...Platform User Manual ...
Страница 33: ......
Страница 41: ......
Страница 96: ...Instrument Operation 63 A test which fails the OBC will appear in the Invalid filter of the Results History ...
Страница 109: ...76 Ancillary Devices LumiraDx Desktop Stand ...
Страница 117: ...84 ...
Страница 118: ......