62 - Instrument Operation
4.10 Onboard Quality Control
In addition to the system checks performed by
the Instrument during start up and within the
test procedure, some LumiraDx Tests also run an
onboard quality control assay alongside the
patient test.
When a test has completed and the OBC test
has passed, the test result will be displayed as
normal. Tapping on the on the result screen at
the end of the test or from the Result History view
shows additional details about the test, including
‘Control: Valid’ which confirms that the OBC has
passed.
If the OBC fails, then an error message will be
displayed and no test result will be returned.
Follow the on screen instructions to dispose of
the Test Strip and start a new test. If the problem
persists, contact Customer Services.