Remove seal
Soak Swab
Squeeze Swab
Attach Dropper Lid
Remove the seal from the top of the
Extraction Vial
containing the
Extraction Buffer
.
Place and soak the
Patient Swab
in the
Extraction Buffer
for 10 seconds then stir well by
rotating the swab against the side of the vial 5
times.
Remove the
Patient
Swab while
squeezing the E
xtraction Vial
to
remove the liquid from the swab.
Discard the swab in biohazard waste.
Firmly attach
the Dropper Lid
to the top of the
Extraction Vial.
The extracted sample must be used (see Step 5 and 6 below) within 5h of
preparation when stored at room temperature.
Running the Test
4. Select the appropriate sample type and
confirm the test type.
5. Gently invert the
Extraction Vial
five times
just before applying the
sample to the
Test Strip
.
6. Apply
one whole drop
of
the sample onto the
Test
Strip Sample Application
Area
when prompted by the
Instrument
.
7. Close the door when
prompted to continue the
test.
8. Results are displayed within 12 minutes of applying the
sample. The left-hand image here shows a positive result for
SARS-CoV-2 Ag and the right-hand image shows a negative
result for SARS-CoV-2 Ag. Tap
Finish
to complete testing or tap
Comment
to leave a comment or to reject the Test, then follow
prompts to return to the
Home Screen
. All test results must be
read using the Lumiradx Instrument.
1. Select
Patient Test
from the
Instrument
Home
Screen
and enter patient details using the
Keyboard
or
Barcode Scanner.
See section 10 of the
Platform User Manual
for
instructions on using the
Barcode Scanner.
2. Remove the
Test Strip
from its
pouch and hold by gripping only the
blue portion.
Do not bend the Test
Strip or touch any part other than the
blue portion.
3. When prompted, open the
Instrument
door and gently insert the
Test Strip
as far as it will go. The thick
black alignment rib on the
Test Strip
should be on the left and line up with the black line on the
Instrument.
Do not apply the sample until prompted.
Install the Lot Calibration file if using a new
Test Strip Lot
for the first time. See section 2.8 of the
Platform
User Manual.
Содержание SARS-CoV-2
Страница 31: ......
Страница 32: ...Platform User Manual ...
Страница 33: ......
Страница 41: ......
Страница 96: ...Instrument Operation 63 A test which fails the OBC will appear in the Invalid filter of the Results History ...
Страница 109: ...76 Ancillary Devices LumiraDx Desktop Stand ...
Страница 117: ...84 ...
Страница 118: ......