13. Review QC result
When the QC test is complete the result will be
displayed on the screen. An example result is
shown here, which shows that the correct result
has been obtained for all assays for this QC
level.
Review the result then tap ‘Finish’ or tap
‘Comment’ to add a comment, if required. A
comment may be up to 50 characters in length.
Once the desired comment(s) are entered tap
‘Done’ to return the result screen.
Tap ‘Finish’ to save the test result to Result History
and complete the test.
14. Remove and dispose of Test Strip
The Instrument will prompt to remove the Test
Strip.
Gently pull the blue label end of the Test Strip to
remove and dispose.
Close the Instrument door.
Dispose of all Test Strips used for
Quality Control testing safely in
accordance with local regulations
and procedures.
Instrument Operation - 55