Example alert message:
Description and instruction
Tap ‘OK’ to proceed. Connect the power supply
unit before the battery runs out.
Example error message: Error 004
Description and instruction
Sample has been applied too early so the test
cannot be completed.
Dispose of Test Strip, start a new test and follow
the instructions on the display.
Example error message: Error 002
Description and instruction
The Instrument door has been opened while
running a test so the test cannot be completed.
Dispose of Test Strip, start a new test and follow
the instructions on the display.
Example error message: Error 006
Description and instruction
When a test is performed the Instrument must
be level and stationary.
Place the Instrument on a level, stable surface
and start a new test.
68 - Troubleshooting