Table 3: Cautions
Warning sign Warning
Do not unplug the sensor cable connectors when the power is on.
Handle the calibration mirror with care.
Do not touch the calibration mirror surface (see
Cleaning the Calibration Mirror
on page 78).
Warning sign Warning
This equipment is not intended for use in residential environments and may not provide adequate
protection to radio reception in such environments. EN 61326‒1:2013.
Notices
This machine must be operated, maintained and repaired only by persons who are properly trained and qualified.
Make sure unauthorized persons cannot operate the machine.
The machine must only be operated by one person at a time.
The machine table moves during operation: ensure a safe area around the machine, a minimum of 0.5 m
(20"). Keep the measurement table surroundings clear of items in order not to block the table movement.
Clear the measurement table of any loose and unnecessary items before operating the machine.
Check the performance of the Emergency Stop button regularly.
Use this machine only for its intended purpose.
Operate this machine only in environments designed and permitted for its use.
Maintenance and service operations must be done regularly and in due time.
Any changes to the safety circuit are strictly forbidden. Changes may cause unexpected start-up or shutdown, or
automatic repetition of cycles, which may lead to serious injuries.
If damage or malfunction occur, shut down the system without delay. Correct the issue immediately. If correcting
the issue is not possible, turn off the system and make sure it cannot be started up before the issue is fixed.
Check that all the power supplies and wires are in good condition. Do not operate the machine if any cracks, worn out
cables or connections are detected.
Make sure all cables are in place and arranged so that there is no risk of tripping.
1.2 Measurement Table
It is recommended to install the UULA measurement system on a stone table or firm worktable, which is suitable for
the scanner weight (120 kg / 265 lbs) and does not wobble underneath the scanner. It is advisable to:
1.
Level the table horizontally.
2.
Set the scanner in place.
3.
Level the scanner using the adjustable legs.
The scanner dimensions are 600 * 880 * 950 mm / 24 * 34.6 * 37.4 inches (D*W*H).
The measurement operation requires using the computer, and a good view of the measuring table. Thus, convenient
access to both the scanner and computer is needed. It is recommended to place both the computer and the scanner on
the same table, which is at least 160 cm (63”) wide.
Содержание FocalSpec
Страница 1: ...FocalSpec 3D Line Confocal Scanner UULA...
Страница 58: ...58 Figure 58 Output Settings 1 Figure 590 Output Settings 2...
Страница 64: ...64 Figure 65 Banding 2 Zooming You can zoom in to view more details Figure 66 Banding Period 300 m...
Страница 67: ...67 Figure 68 Rotating a Picture To open the 3D view select Studies View group 3D View Figure 69 Rotated Picture...
Страница 73: ...73 Figure 77 Rotating a Picture To open the 3D view select Studies View group 3D View Figure 78 Rotated Picture...
Страница 82: ...82 2 11 Declaration of Conformity Figure 92 Declaration of Conformity...