Figure 58. Scanning is on-going in X-direction
You can wait until the whole z-range is measured, or stop the calibration if smaller z-range is enough.
When the scanning is competed turn the sample 90 degrees to Y-direction position like in the picture
below.
Figure 51. The calibration block in Y-Direction scan position.
Press ‘Start Y-Direction’ to continue calibration procedure. Following preview image is visible during
scanning.