MODBUS Register Map v1.49
248083401-1 Rev 5
A-15
Enable Alarming for a Channel
To enable alarming on the third particle channel, the user would enable
Bit 1 for register 43014.
To disable alarming on the third channel and enable alarming on the
second channel, disable Bit 1 for register 43014 and enable Bit 1 for
register 43012.
To disable alarming completely, disable Bit 1 for register 43012. Now,
no channels are enabled for alarms.
Use the Threshold registers to set the alarm threshold value. This is
described in the next section.
43045
unsigned int
Enable for Analog Channel 3 [high]
43046
unsigned int
Enable for Analog Channel 3 [low]
43047
unsigned int
Enable for Analog Channel 4 [high]
43048
unsigned int
Enable for Analog Channel 4 [low]
Table A-14 Alarm Enable Registers
Register Data
Type
Description
Table A-15 Example of Alarming on Channel 2
Registers
Particle
Channel
Bit 1
Enabled
43009 - 43010
1
0
43011 - 43012
2
1
43013 - 43014
3
0
43015 - 43016
4
0
43017 - 43018
5
0
43019 - 43020
6
0
43021 - 43022
7
0
43023 - 43024
8
0
Содержание REMOTE Active Count
Страница 1: ......
Страница 2: ...00...
Страница 6: ......
Страница 10: ...Lighthouse REMOTE Active Count Installation Guide ii 248083401 1 Rev 5...
Страница 12: ...Lighthouse REMOTE Active Count Operating Manual 1 2 248083401 1 Rev 5...
Страница 56: ...MODBUS Register Map v1 49 248083401 1 Rev 5 A 18...
Страница 58: ...Lighthouse REMOTE Active Count Installation Guide B 2 248083401 1 Rev 5...
Страница 61: ...00...
Страница 62: ......