Lighthouse REMOTE Active Count Installation Guide
3-20
248083401-1 Rev 5
Operate the
RAC
Once the RAC is installed and configured, apply power to the
instrument by pressing the bottom Rocker Switch to ON (
|
) position.
The green Power LED will light up. See Figure 3-24.
Figure 3-24 Green Power LED
Note:
LMS Express RT
only measures the
estimated volume of
sample for the Remote
Active Count.
Begin a sampling cycle by using LMS Express RT or by pressing the
Start/Stop button. The yellow Flow LED and the blue Sampling LED
should light up as shown in Figure 3-25.
Figure 3-25 Flow & Sampling LEDs
When the cycle is completed, the pump will shut off and the Flow and
Sampling LEDs will turn off. If the flow does not measure 1 CFM ±
10% the Flow LED will blink until the correct flow is achieved.