MODBUS Register Map v1.49
248083401-1 Rev 5
A-3
40020
ASCII string
Model Name char[10], char [11]
40021
ASCII string
Model Name char[12], char [13]
40022
ASCII string
Model Name char[14], char [15]
40023
unsigned integer
Flow Rate. Divide by 100 to get rate in CFM. For example,
100 = 1CFM.
40024
unsigned integer
Record Count. Total number of records stored in the counter.
40025
unsigned integer
Record Index. Zero based index to data in 30xxx register
series. Must be lower than the record count (register 40024).
Set this index to expose a counter’s record in the 30xxx
registers. Set to -1 to retrieve last record stored in the counter.
40026
unsigned integer
Location number.
Specifies the location of the instrument. Must be 1 to 200 (maps
to location names associated with registers 40200 - 40
9).
40027
signed integer
Real Time Clock (RTC) [high]. Updates instrument’s real-time
clock. Works in conjunction with 40028. Displays date and
time, in number of seconds since midnight, 1/1/1970. Can be
generated by ANSI C/C++ time() function.
40028
signed integer
Real Time Clock [low]
40029
unsigned integer
Initial Delay [high]. Works in conjunction with 40030.
Number of seconds to wait before starting the first sample. Max
value is 86,399, which equals 23h 59m 59s.
40030
unsigned integer
Initial Delay [low]
40031
unsigned integer
Hold Time [high]. Works in conjunction with 40032. Number
of seconds to wait between sample periods. Max value is
86,399, which equals 23h 59m 59s.
40032
unsigned integer
Hold Time [low]
40033
unsigned integer
Sample Time [high]. Works in conjunction with 40034.
Number of seconds to sample. Max value is 86,399, which
equals 23h 59m 59s.
40034
unsigned integer
Sample Time [low]
40035
unsigned integer
Data Set [high]. Works in conjunction with 40036. Data entered
here is applied to the device through the command register.
Table A-3 Sensor Settings Registers
Register Data
Type
Description
Содержание REMOTE Active Count
Страница 1: ......
Страница 2: ...00...
Страница 6: ......
Страница 10: ...Lighthouse REMOTE Active Count Installation Guide ii 248083401 1 Rev 5...
Страница 12: ...Lighthouse REMOTE Active Count Operating Manual 1 2 248083401 1 Rev 5...
Страница 56: ...MODBUS Register Map v1 49 248083401 1 Rev 5 A 18...
Страница 58: ...Lighthouse REMOTE Active Count Installation Guide B 2 248083401 1 Rev 5...
Страница 61: ...00...
Страница 62: ......