Lighthouse REMOTE Active Count Installation Guide
3-10
248083401-1 Rev 5
Install and Set
Up
RAC Pump Instrument
1. Verify that the location where the REMOTE Active Count (RAC)
will be installed is ready for installation:
•
The instrument mounts level on a flat vertical surface;
•
Cabling should be installed and ready;
•
AC power must be available within 3 feet of the mounting
location.
2. Ensure that the RAC is powered OFF (switch is in OFF position).
3. Remove one DIP Switch Cover Plate screw, loosen the other, swing
the cover out of the way and tighten the second screw. Verify DIP
SW 1 and 6 are ON and 7 and 8 are OFF for MODBUS mode
(default setting). If an address other than ’1’ is required, set DIP
SW 1 through 5 to the desired address, up to 31.
Figure 3-13 DIP Switch Positions
4. Return the cover plate to its original position, install removed screw
and tighten both screws.
Note:
Switch 8 controls
the option for Fast Update.
The default setting is ON.
See Table 3-1 on page 13 and Table 3-2 on page 14 for DIP Switch
definitions and addressing.
Note:
Optional Impactor
Assemblies are available
from Lighthouse. Contact a
sales representative at
(510) 438-0500 for more
information.
5. Mount the RAC then remove any protective devices. Refer to the
Read Me First and assemble the Impactor Assembly. Install the
Impactor Assembly on the RAC to complete the following flow
settings steps. See Figure 3-16.
ON
OFF
Содержание REMOTE Active Count
Страница 1: ......
Страница 2: ...00...
Страница 6: ......
Страница 10: ...Lighthouse REMOTE Active Count Installation Guide ii 248083401 1 Rev 5...
Страница 12: ...Lighthouse REMOTE Active Count Operating Manual 1 2 248083401 1 Rev 5...
Страница 56: ...MODBUS Register Map v1 49 248083401 1 Rev 5 A 18...
Страница 58: ...Lighthouse REMOTE Active Count Installation Guide B 2 248083401 1 Rev 5...
Страница 61: ...00...
Страница 62: ......