Get Started
248083401-1 Rev 5
3-13
13. When the LED stops blinking, the settings have been saved and the
pump will stop.
14. Press the start button and run one sample. Verify that the flow reads
28.3. Turn the power OFF when the sample is complete. Remove
the Flow Adjust Cover, tubing and Flow Meter.
Set RAC
Switches
DIP Switch General Definitions
OFF (LEFT) = 0, ON (RIGHT)= 1 as shown in Figure 3-13.
Note:
Use a tool with a
very small pointed tip to
change the DIP Switch
positions.
Table 3-1 DIP Switch Setting Definitions
Position #
Description
Setting
1
Binary Bit 0
Addressing, OFF=0, ON=1
2
Binary Bit 1
Addressing, OFF=0, ON=1
3
Binary Bit 2
Addressing, OFF=0, ON=1
4
Binary Bit 3
Addressing, OFF=0, ON=1
5
Binary Bit 4
Addressing, OFF=0, ON=1
6
Communication
Mode
Always on for MODBUS ASCII
communication
7
Communication
Mode
Always OFF for proper operation
8
Communication
Mode
Fast Update - When ON, records
are not saved to the buffer