Chapter 1: Introduction
7
1
The first line under the meter serial number
indicates when QC is due.
2
The second line indicates when an upload is due.
If the meter is stand-alone, or if configured by a
workstation but no upload lockout was set, this
line is blank.
3
The third line displays
Last upload passed/
failed:
for meters configured by a workstation.
If the meter is in stand-alone mode,
DataLink
inactive
is displayed.
4
The fourth line indicates the date and time of
the last upload. If the meter is in stand-alone
mode, this line is blank.
5
The fifth line displays a battery bar which gives
the approximate status of battery power. If there
is insufficient power to perform a test, the meter
displays a warning message.
The current date and time are displayed at the bottom
of the screen. If the meter was configured by a
workstation, the date and time were downloaded from
the workstation. If the meter is in stand-alone mode,
the date and time were set prior to shipping the meter.
You can change the date and time settings for stand-
alone meters from the Setup screen (see “Setting the
Date and Time” on page 71).
Press
Cont
to advance to the Main Menu. Or, after a
few seconds the meter automatically advances to the
Main Menu. To access the status screen again, you
must turn the meter off, then on again.
Stealth.book Page 7 Monday, October 28, 2002 11:20 PM
Содержание SureStepFlexx
Страница 1: ......
Страница 2: ......
Страница 8: ...vi...
Страница 13: ...1 Introduction...
Страница 28: ...16 SureStepFlexx Meter Operator s Guide...
Страница 29: ...2 Quality Control Test...
Страница 41: ...3 Patient Test...
Страница 51: ...4 Data Transfer and Management...
Страница 61: ...5 Special Tests...
Страница 74: ...62 SureStepFlexx Meter Operator s Guide...
Страница 75: ...6 Care and Maintenance...
Страница 89: ...7 Troubleshooting...
Страница 108: ...96 SureStepFlexx Meter Operator s Guide...
Страница 116: ...104 SureStepFlexx Meter Operator s Guide...
Страница 117: ......