SureStepFlexx Meter Operator’s Guide
If you cannot resolve a problem, call LifeScan Healthcare Professional Line:
1 800 524-7226 (USA)
1 888 353-0800 (Canada)
Shake vial and
apply (High,
Normal, Low)
control
Meter is ready for
control solution
application.
Gently shake vial. Touch a
drop of control solution to
center of pink test square.
• If white pad becomes sat-
urated, you have applied
too much solution. Repeat
application with new test
strip.
• If white patches or streaks
appear on confirmation
dot, you have not applied
enough solution. Repeat
application with new test
strip.
Strip Lot
Expired
Test strip lot expired.
Enter and use new test strip
lot. Discard any remaining
expired test strips.
Upload
Overdue
Data transfer not
performed within
specified time.
Perform data transfer session.
Wrong Control
Type
Type of control
(solution) entered does
not match test type.
Check control type (low,
normal, high, linearity, or
unknown) and reenter
information.
MESSAGE
POSSIBLE CAUSE
SOLUTION
Stealth.book Page 90 Monday, October 28, 2002 11:20 PM
Содержание SureStepFlexx
Страница 1: ......
Страница 2: ......
Страница 8: ...vi...
Страница 13: ...1 Introduction...
Страница 28: ...16 SureStepFlexx Meter Operator s Guide...
Страница 29: ...2 Quality Control Test...
Страница 41: ...3 Patient Test...
Страница 51: ...4 Data Transfer and Management...
Страница 61: ...5 Special Tests...
Страница 74: ...62 SureStepFlexx Meter Operator s Guide...
Страница 75: ...6 Care and Maintenance...
Страница 89: ...7 Troubleshooting...
Страница 108: ...96 SureStepFlexx Meter Operator s Guide...
Страница 116: ...104 SureStepFlexx Meter Operator s Guide...
Страница 117: ......