Chapter 5: Special Tests
55
◆
IMPORTANT:
You have up to 2 minutes to insert the
test strip after the linearity solution is applied. If
you insert the test strip after 2 minutes, you may get
an out-of-range result or an error message. Discard
the test strip and repeat the test with a new test strip.
The test result appears in
approximately 30 seconds.
The time may vary depending
on the temperature and glucose
level.
◆
IMPORTANT:
Do not remove
the test strip until the
countdown is complete.
■
NOTE:
The meter can display linearity results up to
766 mg/dL.
10
Press
to display the
Linearity Level screen. Press
Run REP 2
to proceed with
the remaining replicates of
level 1.
Or, if you wish to record
additional information with the
test result, press
Entr Note
and
choose 1 to 3 comments from the list. Press
.
■
NOTE:
Results tagged with the comment “Procedure
Err” are not included in the statistical calculations
for any linearity tests.
11
When you complete all replicates for level 1,
press
Next Level
from the Linearity Level screen
to proceed with the remaining linearity solution
levels (2 through 5).
L i n e a r i t y 1 R e s u l t
45
mg/dL
Entr
Note
L i n e a r i t y 1 L e v e l
R u n
Next Level
R E P 2
Stealth.book Page 55 Monday, October 28, 2002 11:20 PM
Содержание SureStepFlexx
Страница 1: ......
Страница 2: ......
Страница 8: ...vi...
Страница 13: ...1 Introduction...
Страница 28: ...16 SureStepFlexx Meter Operator s Guide...
Страница 29: ...2 Quality Control Test...
Страница 41: ...3 Patient Test...
Страница 51: ...4 Data Transfer and Management...
Страница 61: ...5 Special Tests...
Страница 74: ...62 SureStepFlexx Meter Operator s Guide...
Страница 75: ...6 Care and Maintenance...
Страница 89: ...7 Troubleshooting...
Страница 108: ...96 SureStepFlexx Meter Operator s Guide...
Страница 116: ...104 SureStepFlexx Meter Operator s Guide...
Страница 117: ......