Chapter 5: Special Tests
61
In-Service Test
Use the In-Service mode to train and certify operators.
This mode is identical to the QC Test mode. However,
in-service test results cannot be viewed in Review
Data mode and appear only on the Data Log Report.
■
NOTE:
If the meter is set for Meter lockout (see QC
enforcement on page 19), and QC is overdue, you will
not be able to access Special Tests until QC testing is
successfully performed.
1
Select
Special Tests
from the
Main Menu.
2
Select
In-Service Mode
from
the Special Tests menu.
3
Follow the instructions in Chapter 2, Quality
Control Test, beginning with step 4 on page 22.
Use the SureStepPro Glucose Control Solution
levels as required by your institution.
S e l e c t S p e c i a l T e s t
L i n e a r i t y T e s t s
U n k n o w n T e s t s
I n - S e r v i c e M o d e
Stealth.book Page 61 Monday, October 28, 2002 11:20 PM