66
SureStepFlexx Meter Operator’s Guide
swab dampened with water to remove residual
bleach.
▲
CAUTION:
Bleach residue left on the test strip holder
may lead to an error message or an inaccurate, high
result.
3
Dry the test strip holder with a soft cloth or
lint-free tissue. Close the holder and set it aside.
4
Using a cotton swab or soft cloth dampened
with a 10% bleach solution, wipe the lens area
and contact points. Wipe this area even if it
doesn’t appear to be dirty. Use a cloth moistened
with water to remove residual bleach.
Be careful not to scratch the lens area or get water
inside the meter.
contact points
lens area
Stealth.book Page 66 Monday, October 28, 2002 11:20 PM