![LI-COR LI7200RS Скачать руководство пользователя страница 46](http://html.mh-extra.com/html/li-cor/li7200rs/li7200rs_instruction-manual_1906669046.webp)
Section 3. Operation
Select the communication baud rate for
the serial port. The maximum update
rate is dependent upon the rate available
with your computer's serial port and the
Update Rate
used while the program
communicates with the instrument.
Select from 9600, 19200, 38400, 57600,
or 115200.
The Update Rate is the communication
frequency between the instrument and
the computer. Select from 0.1, 0.2, 0.5, 1,
2, 5, 10, or 20 Hz.
Note that at 9600
baud, the maximum update frequency is
2 Hz; at 19200 baud, 5 Hz; at 38400
baud, 10 Hz; at 57600 baud, 15 Hz; and
at 115200 baud, 20 Hz.
Click the
Con-
nect
button to establish communications
with the instrument.
Run Disconnected
3-4
LI-7200RS Enclosed CO₂/H₂O Analyzer
Содержание LI7200RS
Страница 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Страница 2: ......
Страница 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Страница 12: ...x...
Страница 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Страница 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Страница 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Страница 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 225: ......