Section 4. User Calibration
Zero CO
2
IMPORTANT:
Always zero the instrument before spanning (don’t span, then
zero).
5
In the PC software, click
Calibration
under the LI-7200 button.
Verify that temperature and pressure sensors are working properly by checking
their values in the dashboard.
6
Click on the
Manual
tab and view the value of
Z
co
(CO
2
Zero).
7
Click the
Zero
tab.
When the reading has stabilized in the dashboard, and the Gas Stability flag is
green, click
Zero CO2
to set the CO
2
zero.
After a brief delay, the displayed CO
2
value should be fluctuating around zero.
Check the value of
Z
co
shown on the
Manual
tab. It should be near 1 (typically
between 0.85 and 1.1). This value will steadily increase over time (2-3 months) as
the internal chemicals lose effectiveness.
4-6
LI-7200RS Enclosed CO₂/H₂O Analyzer
Содержание LI7200RS
Страница 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Страница 2: ......
Страница 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Страница 12: ...x...
Страница 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Страница 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Страница 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Страница 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 225: ......