1
Choose a source to be integrated from the Integration Channel list.
2
Choose the method to start the integration; immediately, or using a threshold value.
3
Choose the method to stop the integration; manually, with a threshold value, or after a user-
entered elapsed time (s) has expired.
4
Enter the threshold value for the start or stop (or both) of the integration, if either was chosen in
Steps 2 or 3 above.
5
Click on the Apply button.
If Start Immediately was chosen for the Start Time, the integration function is star-
ted and/or stopped manually by clicking the
∫
dt Integration On
or
∫
dt Integration Off
buttons (available on the dashboard, as well).
Example:
Start integrating CO
2
(µmol/mol) after it reaches a value of 500 µmol
mol
-1
, and continue integrating for 5 minutes.
6
Choose
CO2 (µmol/m
3
)
from the Integration Channel list.
7
Click the
Start when channel reaches
radio button and enter 500 in the text entry field.
8
Click the
Stop after an elapsed time
radio button and enter 300 (seconds) in the text entry
field.
9
Click
Apply
.
10
Click the
∫
dt Integration On
button. Click
OK
to dismiss the dialog.
The integration result (area under the curve) can be viewed in the dashboard.
Section 7. Software Reference
7-9
Integration
Содержание LI7200RS
Страница 1: ...Enclosed CO2 H2O Gas Analyzer Instruction Manual...
Страница 2: ......
Страница 11: ...ix Appendix C Suppliers Appendix D Configuration Grammar Appendix E Warranty Appendix F Index...
Страница 12: ...x...
Страница 20: ...Section 1 General Information 1 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 54: ...Section 3 Operation 3 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 112: ...Section 4 User Calibration 4 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 130: ...Section 5 Maintenance 5 18 LI 7200RS Enclosed CO H O Analyzer...
Страница 138: ...Section 6 Troubleshooting 6 8 LI 7200RS Enclosed CO H O Analyzer...
Страница 184: ...Section 8 Theory of Operation 8 12 LI 7200RS Enclosed CO H O Analyzer...
Страница 187: ...Gas Analyzer Dimensions Appendix A Specifications A 3 Appendix A Specifications...
Страница 188: ...Appendix A Specifications LI 7550 Dimensions A 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 192: ...Appendix B Pin Assignments B 2 LI 7200RS Enclosed CO H O Analyzer...
Страница 196: ...Appendix C Suppliers C 4 LI 7200RS Enclosed CO H O Analyzer...
Страница 225: ......