4-8
A new offset term
S
c0
is then computed using equation 4-5, since the slope term
S
c1
has changed. Similarly, for H
2
O, each time a normal water span is set, the
instrument retains
4-9
and when a secondary H
2
O span is performed at water mole fraction
W’
f
with meas-
ured absorptance
α
'
w
, then the span slope term is computed from
4-10
and a new offset term
S
w0
is computed from equation 4-6.
Section 4. User Calibration
4-11
What Actually Happens