9
Turn the Fluid Warmer ON. Let the solution circulate
for a 30-minute disinfection period.
10
Turn the Fluid Warmer OFF.
11
Empty the reservoir.
12
Remove the Disposable Set and discard according to established
hospital procedures
Change Recirculating Solution with a 0.3% Hydrogen
Peroxide/Distilled Water Solution
1
Place a container under the drain valve of the Fluid Warmer.
2
Drain the recirculating solution by turning the drain valve
clockwise 90 degrees.
3
When all the solution has drained from the reservoir, close the
drain valve.
4
Prepare a 0.3% hydrogen peroxide/distilled water solution. Mix
140 ml of 3% hydrogen peroxide solution and 1,260 ml of
distilled water.
5
Fill the reservoir with 0.3% hydrogen peroxide/distilled water
solution. The reservoir holds 1.4 liters.
Change O-Rings
Change the O-Rings in the #1 Block and #2 Block.
1
Remove each O-Ring from its socket by pulling it out
with a pair of needle-nose pliers or by prying it out with
a small screwdriver.
2
Lubricate the new O-Rings from the O-Ring Kit.
3
Press each O-Ring into its socket.
Clean Fan Filter
The Fan Filter (
a
) is located on the bottom of the Fluid Warmer.
The fan guard snaps in place.
1
Remove four screws and unsnap the fan guard from the bottom
of the unit.
2
Clean the filter with warm soapy water.
3
Replace the fan guard and filter.
S E C T I O N 1 2 • M a i n t e n a n c e
|
Operator’s Manual
62
a
Содержание H-1200 115V
Страница 1: ...OPERATOR S MANUAL P N 4533708GB Rev 006 lTM s...
Страница 2: ...ii Operator s Manual...
Страница 24: ...S E C T I O N 6 P r i n c i p l e o f O p e r a t i o n Operator s Manual H 1200 H 1200 18...
Страница 33: ...Operator s Manual S E C T I O N 7 C o n t r o l s a n d D i s p l a y s 27...
Страница 34: ...Operator s Manual 28 S E C T I O N 8 O p e r a t i o n...
Страница 78: ...Operator s Manual S E C T I O N 1 6 Symbols S E C T I O N 1 6 S y m b o l s 72...
Страница 79: ...S E C T I O N 1 6 S y m b o l s Operator s Manual 73...