48
S E C T I O N 9 • O p e r a t i n g I n s t r u c t i o n s
9.3 Replace the Gas Vent/Filter Assembly
Use one of the following Gas Vent/Filters Assemblies:
• F-10 Gas Vent / Filter Assembly for DI-50, D/DI-60HL, D/DI-70,
and D/DI-100 Disposable Sets
• F-30 Gas Vent / Filter Assembly for D/DI-300 Disposable Sets
• F-50 Gas Vent / Filter Assembly for DI-55, D/DI-65HL, D/DI-75,
D/DI-150, and D/DI-350 Disposable Sets
1
Close all clamps above and below the Gas Vent/Filter Assembly on
D/DI-series Disposable Set and on new Gas Vent/Filter Assembly.
2
Turn the Fluid Warmer OFF.
Note:
Steps 3 and 4 are different for the H-1200 and H-1000 Fluid
Warmers. Continue with the appropriate steps 3 and 4 that apply for
your specific Fluid Warmer.
3
Remove the
used
Gas Vent/Filter Assembly while still connected to
the Disposable Set:
a
Remove the Gas Vent/Filter Assembly from the #4 Block.
For DI-55, D/DI-65HL, D/DI-75, D/DI-150, and D/DI-350
disconnect the clip from the bottom of the Heat Exchanger.
b
Open the #3 Clamp Slot door and remove the patient line from
the Clamp Slot.
4
Install the
new
Gas Vent/Filter Assembly:
Note:
The steps to install the Gas Vent/Filter Assembly differ
depending on the Disposable Set. Continue with the appropriate
steps that apply for your Disposable Set.
Use the following steps for DI-55, D/DI-65HL, D/DI-75,
D/DI-150, and D/DI-350.
a
Press the
new
Gas Vent/Filter Assembly into the #4 Block and
press the clip to the bottom of the Heat Exchanger.
b
Insert patient line in the Clamp Slot.
c
Align the patient line in the Clamp Slot without kinking.
d
Close the Clamp Slot door.
e
Go to step 5 to continue.
Use the following steps for DI-50, D/DI-60HL, D/DI-70,
D/DI-100, and D/DI-300.
a
Insert patient line in the Clamp Slot.
b
Close the Clamp Slot door.
c
Align the patient line in the Clamp Slot without kinking.
d
Press the Gas Vent/Filter Assembly into the #4 Block.
e
Go to step 5 to continue.
|
Operator’s Manual
F-10/30
F-50
Содержание H-1200 115V
Страница 1: ...OPERATOR S MANUAL P N 4533708GB Rev 006 lTM s...
Страница 2: ...ii Operator s Manual...
Страница 24: ...S E C T I O N 6 P r i n c i p l e o f O p e r a t i o n Operator s Manual H 1200 H 1200 18...
Страница 33: ...Operator s Manual S E C T I O N 7 C o n t r o l s a n d D i s p l a y s 27...
Страница 34: ...Operator s Manual 28 S E C T I O N 8 O p e r a t i o n...
Страница 78: ...Operator s Manual S E C T I O N 1 6 Symbols S E C T I O N 1 6 S y m b o l s 72...
Страница 79: ...S E C T I O N 1 6 S y m b o l s Operator s Manual 73...